Table 4 Comparison of αSyn_RT-QuIC results between ITA-lab and USA-lab

From: Discrimination of MSA-P and MSA-C by RT-QuIC analysis of olfactory mucosa: the first assessment of assay reproducibility between two specialized laboratories

 

PD

MSA-P

MSA-C

HS

ITA-lab

USA-lab

ITA-lab

USA-lab

ITA-lab

USA-lab

ITA-lab

USA-lab

Samples analyzed

13

13

20

20

10

10

11

11

αSynD seeding activity

9

9

18

18

0

1

0

1

Mean fluorescence at the time threshold % (SD)

56.46 (39.88)

41.85 (42.41)

56.70 (30.76)

46.10 (32.20)

5.90 (0.99)

11.00 (19.40)

6.18 (1.07)

10.00 (19.29)

Mean time at fluorescence threshold (SD)

9.31 (1.19)

15.93 (4.41)

10.54 (1.17)

17.23 (3.29)

NA

NA

NA

NA

Seeding activity vs clinical diagnosis (sensitivity)

69%

69%

90%

90%

0%

10%

NA

NA

Disease-specific interrater agreement between labs

100%

100%

90%

91%

Overall interrater agreement between labs

96% (Kappa = 0.93, 95% CI 0.83–1.00)